Product-Specific Commands—Description
:INITiate:CONTinuous
Description
This command is used to select whether the
trigger system is continuously initiated or not. The
trigger system is used to control trace acquisition.
At *RST, this value is set to off (disabled).
Syntax
:INITiate:CONTinuous<wsp><Continuous>
Parameter(s)
Continuous:
The program data syntax for <Continuous> is
defined as a <Boolean Program Data> element.
The <Continuous> special forms ON and OFF are
accepted on input for increased readability. ON
corresponds to 1 and OFF corresponds to 0.
The <Continuous> parameter corresponds to the
new state of the trigger system continuous cycle.
0 or OFF: disable continuous cycle: trigger
system returns to idle.
1 or ON: enable continuous cycle.